How It Works

Study Design

Short-term and long-term precision are routinely assessed in both verification and validation. Choose the design that matches the sources of variability you need to characterize:

Simple PrecisionShort-term / within-runTEa-based or manufacturer claimSingle run
PurposeRapid assessment of within-run repeatability
Run structureSingle run or continuous run
ReplicatesTypically 10–20 replicates in one run
Concentration levelsMultiple levels recommended (e.g. low and high); each level evaluated independently
Precision typeRepeatability (within-run) only — does not capture between-run or between-day variation
ComparisonObserved SD vs. Goal SD (derived from TEa) or manufacturer claim

Because all replicates are collected in a single run, Simple Precision measures only within-run repeatability. Use a Complex (multi-day) design when between-run or between-day sources of variability must be estimated.

Complex Precision — Short-termShort-term multi-dayCLSI EP15-A35 × 1 × 5
Days5 minimum
Runs per day1 run per day
Replicates per run5 replicates per run
Concentration levels2 levels recommended (low and high)
Total measurements25 per concentration level (minimum)

The 5 × 1 × 5 design yields a degrees-of-freedom for repeatability (dfR) of 20 — the minimum EP15-A3 requires for a statistically valid short-term assessment. Calculates within-run (repeatability) and within-laboratory (total) precision. Used in both verification and validation contexts when a short multi-day protocol is appropriate.

Complex Precision — Long-termLong-termCLSI EP05-A320 × 2 × 2
Days20 days
Runs per day2 runs per day
Replicates per run2 replicates per run
Concentration levelsAt least 2 levels recommended
Total measurements80 per concentration level

The 20 × 2 × 2 design captures within-run, between-run, between-day, and within-laboratory variance components via ANOVA. The two-run-per-day structure is essential for separating between-run from between-day sources of variability. Used in both verification and validation when long-term precision components must be characterized.