Interpreting Your Results

Carryover Analysis Results

The results panel displays the following statistics:

MetricWhat It Tells You
Low (Control) MeanThe mean of the 3 baseline Low specimens (positions 1–3). This is the reference value representing the true low concentration free of carryover contamination.
Low (Control) SDThe standard deviation of the 3 baseline Low specimens. Reflects normal variability in the low-concentration pool and is used in the Auto error limit calculation (3 × SD).
Low (Test) MeanThe mean of the 8 Low test specimens that follow High specimens. Elevated values relative to the control mean suggest carryover contamination from the preceding high-concentration measurements.
Low (Test) SDThe standard deviation of the 8 Low test specimens. Higher variability here than in the control group may indicate inconsistent carryover contamination across the sequence.
Carryover MeanLow (Test) Mean minus Low (Control) Mean. The estimated average carryover contamination. A positive value indicates the test specimens are reading higher than baseline. A value near zero indicates minimal carryover.
Upper Confidence LimitThe upper bound of the 95% CI for the carryover estimate, calculated using the Welch-Satterthwaite degrees of freedom. This is the worst-case carryover with 95% confidence and is the value compared against the Error Limit for pass/fail.
Error LimitThe maximum acceptable carryover as configured (Auto = 3 × Control SD; Concentration = user-defined absolute; Percent = user-defined % of low concentration).
ResultPASS if Upper Confidence Limit < Error Limit; FAIL if Upper Confidence Limit ≥ Error Limit.

Interpreting the Carryover Mean and UCL

The Carryover Mean alone does not determine pass/fail — the Upper Confidence Limit (UCL) does. Even a small Carryover Mean can produce a failing UCL if the variability in the test or control group is large, because the UCL widens with higher uncertainty. Conversely, a modest positive Carryover Mean may still pass if the data are consistent and the UCL remains below the error limit.

When interpreting a passing result, confirm that the UCL has meaningful margin below the error limit. A UCL very close to the error limit — even a passing one — warrants monitoring, particularly before testing high-concentration patient specimens in routine use.