Interpreting Your Results

General Statistics

The General Statistics panel provides a top-level summary of the study:

StatisticWhat It Tells You
Overall Mean (Error Index by Instrument)The mean Error Index across all non-target instrument measurements. A value near 0 indicates that the fleet as a whole is well-centered around the target. A non-zero mean suggests a systematic directional bias across instruments.
Target RangeThe minimum and maximum target values observed across all specimens in the study. Reflects the concentration range over which inter-instrument agreement was evaluated.
Coverage RatioThe percentage of the reportable range (AMR) spanned by the study's target values. Higher coverage indicates that inter-instrument agreement was assessed across a broader portion of the clinically relevant concentration range.
InstrumentsThe total number of instruments included in the study.
SamplesThe total number of unique specimens tested.
Number of Data Points Outside of Allowable ErrorThe count of individual measurements where |Error Index| > 1.0 (FAIL), expressed as a fraction of total non-target measurements. For example, 0/6 means all 6 measurements from non-target instruments fell within allowable error.

Result Summary by Instrument

The Result Summary table lists every measurement from every non-target instrument, with the following columns:

ColumnWhat It Tells You
InstrumentThe instrument name and serial number identifying which analyzer produced this measurement.
Sample IDThe specimen identifier for this row.
ResultThe raw measured value from this instrument for this specimen.
TargetThe reference value for this specimen — either the median of designated target instruments or the median of all instruments if no target is designated.
Observed ErrorResult minus Target. Positive values indicate the instrument reads higher than the target; negative values indicate lower.
TEaThe allowable error window at this specimen's concentration, expressed as ± an absolute value. Derived from the analyte's configured TEa and bias allocation percentage.
Error IndexObserved Error divided by TEa. |Error Index| ≤ 1.0 = PASS; |Error Index| > 1.0 = FAIL. Displayed as '–' for target instruments.
StatusPASS (green) if |Error Index| ≤ 1.0; FAIL (red) if |Error Index| > 1.0. Shown as '–' for target instruments, which are not evaluated.

Error Index by Instrument

This dot plot displays every individual Error Index value, with instruments arranged along the Y-axis and the Error Index on the X-axis (−1.5 to +1.5). Each dot represents one measurement. Pink shaded regions beyond ±1.0 mark the unacceptable error zone.

  • Dots clustered near zero — Instrument results closely match the target across specimens. Good inter-instrument agreement.
  • Dots consistently offset in one direction — The instrument has a systematic positive or negative bias relative to the target. The direction and magnitude suggest a calibration or reagent issue specific to that instrument.
  • Dots in the pink zone (|Error Index| > 1.0) — One or more measurements exceed allowable error. Review the Result Summary table to identify which specimens are affected and investigate root causes.
  • Sample ID Finder — When enabled, hovering over any data point highlights all points from the same specimen across all instruments in green. This makes it easy to determine whether a problematic result is isolated to one instrument or reflects a specimen-level issue (e.g., sample integrity).

Scatter Plot (All Instruments)

The scatter plot displays each instrument's measured value (Y-axis) against the corresponding target value (X-axis). It overlays:

  • Data points — One point per measurement per instrument, color-coded by instrument.
  • Perfect Agreement Line (Y = X) — Represents exact agreement between each instrument and the target. Points on this line indicate zero bias.
  • TEa bounds — Parallel lines above and below the perfect agreement line representing the allowable error window. Points outside these bounds correspond to Error Index magnitudes > 1.0.
  • AMR limits — Vertical markers indicating the lower and upper bounds of the configured reportable range.
  • Instrument Finder — When enabled, hovering over a data point highlights all points from the same instrument in green, making it easier to identify instrument-specific patterns or systematic offsets.

Error Index by Concentration (All Instruments)

This scatter plot shows the Error Index (Y-axis) for each measurement against the corresponding target concentration (X-axis). It includes:

  • Data points — Each point represents one measurement. Points in pink zones (|Error Index| > 1.0) exceed allowable error.
  • Zero line — The horizontal reference line at Error Index = 0, representing perfect agreement with the target.
  • AMR limits — Vertical markers at the reportable range boundaries. Points outside the AMR limits are flagged as out-of-range.
  • Horizontal scatter pattern — A random, flat scatter around zero across concentrations indicates good precision and no concentration-dependent bias.
  • Systematic patterns — A fan-shaped spread (errors increasing with concentration) or a curved trend may indicate proportional bias or a non-linear relationship between the instrument and the target. Investigate calibration and reagent lot consistency.
  • Instrument Finder — When enabled, hovering over a data point highlights all points from the same instrument in green to reveal instrument-specific concentration-dependent patterns.

All three figures support zoom for closer inspection of dense data regions.