How It Works

Study Design

The MIMC module requires each specimen to be measured on all instruments included in the study, enabling direct cross-instrument comparison at the same concentration. The minimum recommended design is:

Multi-Instrument Method ComparisonVerificationCLSI EP31-A-IR2–30 instruments
Instruments2–30 (minimum 2; one may be designated as target)
SpecimensMinimum 3; 5–10 recommended
Replicates per specimenMinimum 1 per specimen per day
DaysMultiple days recommended to assess reproducibility
Concentration rangeSpecimens should span the reportable range
Paired observationsMinimum 3 paired observations required

All instruments must measure the same specimens to enable direct comparison. Concentrations should be distributed across the reportable range — not clustered at a single level — to evaluate agreement across clinically relevant values.